<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>17029</ErrorNumber>
<ErrorFamily>Exposure Head</ErrorFamily>
<ErrorMessage>
HEAD: head controller incompatible software version
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Thermal</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
This error is only reported when the head initializes (ie. on MPE
startup). Note that this error does not cripple the system. Proceed
at your own risk if you see this error. The customer should never see
this error though!
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Call service immediately for a firmware upgrade.
</text></DescriptionText>
</OperatorAction>
<Cause>
The MPE firmware has detected head firmware that it does not believe
it is compatible with. This may cause many unknown problems.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:34 2008 PST
</Updated>
<Links>
</Links>
</Error>